Bayesian errors-in-variables calibration
Research context
research method · conditional · As of 2026-10-10
A calibration likelihood can treat reference values x_i as latent uncertain quantities, with measured standards r_i and instrument indications y_i modelled jointly around x_i and f(x_i,θ).
Key knowledge
A calibration likelihood can treat reference values x_i as latent uncertain quantities, with measured standards r_i and instrument indications y_i modelled jointly around x_i and f(x_i,θ). Posterior inference p(θ,x|r,y) combines the joint likelihood with documented priors; factorizing reference and instrument likelihoods additionally requires conditional independence. Ignoring reference uncertainty can bias slopes and understate inverse-calibration uncertainty.
Values and equations
p(θ,x|r,y)∝p(r,y|x,θ)p(x,θ)
- θ
- Calibration parameters with model-specific units
- x,r
- Latent and measured reference quantities with measurand units
- y
- Instrument indications with instrument units
- p
- Joint/conditional densities on the stated parameter space
Conditions and scope
- Specify reference covariance, instrument noise, units, calibration range and prior information without counting the same certificate twice.
- For a new indication, propagate posterior parameter/reference uncertainty through the inverse, including multiple roots/nonlinearity. A proper posterior alone is not metrological traceability.
Sources · 1
References
Sources retain their recorded scope. A source link alone does not establish verification of every statement.
Evidence and review
Inspected §§8.5, 10.4–10.6, 11.4, 11.6–11.10, 12 and Annex B: shared effects, inverse calibration, time series and prior sensitivity; source check dated 2026-10-10. Scope is limited to the cited statements, equations and assumptions, not the whole publication.
2026-10-10 · Author source inspection of the displayed claim, model assumptions and equations; independent exact-content and branch-inventory review is a separate gate. Contemporary status is a dated snapshot, not a promise of ongoing monitoring.